TECHNICAL DATA

 

 

Displacement

 

     50 Hz operation

2.2 ft3 min-1 (3.7m3h-1)

     60 Hz operation

2.6 ft3 min-1 (4.5m3h-1)

Speed (Pneurop 6602)

 

     50 Hz operation

2 ft3 min-1 (3.3m3h-1)

     60 Hz operation

2.3 ft3 min-1 (3.9m3h-1)

 

 

Ultimate vacuum (total pressure)

Gas ballast

High vacuum mode

High throughput mode

Closed

1.5 x 10-3 torr

2.3 x 10 -3

 

2 x 10-3 mbar

3 x 10-2 mbar

Low flow, I

2.3 x 10-2 torr

4.6 x  10-2 torr

 

3 x 10-2 mbar

6 x 10-2 mbar

High flow, II

 

9.1 x 10-2 torr

 

 

1.2 x 10-1 mbar

 

 

Inlet connection

NW25

 

Outlet connection

NW25

 

Maximum allowed outlet pressure

14 psig (1 bar gauge)

 

Maximum allowed inlet and gas ballast

 

Pressure

7 psig (0.5 bar gauge)

 

Maximum water vapor inlet pressure

60 torr (80 mbar)

 

Maximum water vapor pumping rate

 

     Gas ballast low flow

60 gh-1

     Gas ballast high flow

220 gh-1

 

Motor power

 

     50 Hz

250 W

     60 Hz

300 W

 

Electrical supply

 

     1-phase

110-120 V or

 

220-240 V, 50 or 60 Hz

     3-phase

200-230 V or

 

380-460 V, 50 or 60 Hz

 

Operating temperature range

12 - 40 oC

 

Weight, without oil

47.3 lbs(21.6 kg)

 

Noise

48 dB(A)

 

Oil capacity

 

Maximum

0.71

Minimum

0.421

 

Recommended oil (supplied)

Ultragrade 19